Create a Report

Please provide report title
Please provide location
Please provide details

Get alerts and updates for your case!

This information is not shown on the website. We can notify you of updates, and may contact you for more information to help resolve or follow up on your issue.

Please provide email or phone
Please provide email or phone

We do not show your email or contact info

Skip this step

Add photos or video

We recommend photos and videos to help explain your report

Add Photo/Video
Drop files here
Skip this step

Email or SMS copy of report

Enter below to get emailed a copy of your report, or sms a link to your report

Please provide email or phone
Please provide email or phone

Thank you for your report!

The report was
successfully sent!


Your report is queued - it may take up to 12 hours to process your report.


Email us at support@safelyhq.com if you have questions.

Report by

Safety Report: Infusion System Administration Set (infusion set) - recalled due to Use of the pump system potentially could cause over-infusion or under-infusion of therapy, USA

4 years ago source fda.gov

Recall notice

United States

Company name: CME America
Brand name: BodyGuard
Product recalled: Infusion System Administration Set (infusion set)
Reason of the recall: Use of the pump system potentially could cause over-infusion or under-infusion of therapy
FDA Recall date: July 08, 2020
Recall details: PRNewswire/ - CME America, a wholly owned subsidiary of BD (Becton, Dickinson and Company), announced a voluntary recall for all CME America BodyGuard® Infusion System Administration Sets (infusion sets) used with the company's BodyGuard® infusion pumps—that were distributed beginning May 2016 (see complete impacted product list below).  This action was initiated on June 16, 2020. As part of CME America's commitment to quality, following the previously announced recall (April 27, 2020) of the BodyGuard® Infusion Pump Systems, the company conducted additional flow-rate accuracy testing.

This testing revealed that some infusion sets do not meet the ±5% delivery accuracy for the system or the ±13% accuracy identified in the earlier recall notification (www.bd.com/en-us/support/recallnotifications/recall-notification---cme-america-bodyguard-infusion-pump-system). Therefore, the use of the pump system potentially could cause over-infusion or under-infusion of therapy and patient harm. Based on those test results CME America is providing additional information and customer actions regarding its previous recall to include all infusion sets used with the BodyGuard® infusion pump distributed beginning May 2016.

CME America has defined four categories of impacted infusion sets based on delivery inaccuracy variability, two of which (Category A and B) can continue to be used in accordance with the instructions in the recall letter (www.bd.com/en-us/support/recall-notifications/recall-notification---cmeamericabodyguard-infusion-administration-sets).  CME America also announced on April 27, 2020, the decision to suspend distribution of the BodyGuard® infusion pumps and to remove all existing products from the U.S. market. CME America will work with customers to address the latest expanded infusion set recall and will continue to maintain continuity of care during the COVID-19 pandemic. Until such time that the BodyGuard® pumps have been removed from the market, CMEA will supply accessories and infusion sets in "Category A" and "Category B" to support the infusion pumps, and customers can continue to use the products in accordance with the Operator's Manual and the additional mitigations outlined in the customer letter.

Affected Product Types:

- BodyGuard BodySet
- BodyGuard Microset with Needleless Adaptor
- BodyGuard Microsets
- BodyGuard Microset w/ Non-Vented Spike Connector
- BodyGuard set with Female Luer
- BodyGuard Microset w/ 0.2 mic filter and lower y-site with female luer
- BodyGuard Microsets with Filter
- Standard BodySet with Needleless Connectors
- BodyGuard Microset with Filter and Manual Priming Valve
- CMExpress Microbore Sets
- BodyGuard Microset with Male Luer Connectors
- CMExpress Needleless Y Site Microbore Set

Check the full recall details on www.fda.gov/safety/recalls-market-withdrawals-safety-alerts/cme-america-announces-follow-voluntary-recall-bodyguardr-infusion-system-administration-sets

Source: FDA