Create a Report

Please provide report title
Please provide location
Please provide details

Get alerts and updates for your case!

This information is not shown on the website. We can notify you of updates, and may contact you for more information to help resolve or follow up on your issue.

Please provide email or phone
Please provide email or phone

We do not show your email or contact info

Skip this step

Add photos or video

We recommend photos and videos to help explain your report

Add Photo/Video
Skip this step

Email or SMS copy of report

Enter below to get emailed a copy of your report, or sms a link to your report

Please provide email or phone
Please provide email or phone

Thank you for your report!


IMPORTANT - Your report is queued
It may take up to 12 hours to process your report.


Email us at support@safelyhq.com if you have questions.

Report by

Safety Report: Several Brand CPAP and BIPAP Masks with Magnets - recalled due to interference with medical implants, USA

2 months ago source fda.gov

Recall notice

United States

On March 1, 2024, Sleepnet Corporation initiated a worldwide recall for all CPAP and BIPAP masks with magnets due to potential interference with certain medical devices. When a magnet comes into close proximity to certain medical implants or metallic implants, it could interfere with the performance or the position of the implant, potentially resulting in serious injury or death. Sleepnet has been distributing masks with magnets worldwide since 2006 and to date, there have been no Medical Device Reports associated with the Sleepnet masks with magnets.

Affected products are Mojo Full Face Vented Mask, Mojo Full Face Non‐Vented Mask, Mojo 2 Full Face Vented Mask, Mojo 2 Full Face Non‐ Vented Mask, Mojo 2 Full Face AAV Non‐Vented Mask, iQ 2 Nasal Mask, and Phantom 2 Nasal Mask. This is applicable to all lot/UDI numbers.

Sleepnet masks with magnets are safe when used in accordance with the newly updated Instructions for Use. Continue using masks according to the updated instructions and labeling if patients or people in close proximity to them do not have implanted metallic medical devices or metallic objects in the body.

Sleepnet is adding a new contraindication and an updated warning to the Instructions for Use labeling of the affected products. This will be included in the labeling for all future production of these masks. The labeling will be updated to state the following:
1- New Contraindication:
Do not use this mask if you or anyone (example: household members, bed partners, caregivers, etc.) in close physical contact with your mask has an active medical implant or metallic implant that will interact with magnets. Implant examples include, but are not limited to, pacemakers, implantable cardioverter defibrillators (ICD), neurostimulators, aneurysm clips, metallic stents, ocular implants, insulin/infusion pumps, cerebral spinal fluid (CSF) shunts, embolic coils, metallic splinter, implants to restore hearing or balance with implanted magnets (such as cochlear implants), flow disruption devices, contact lenses with metal, dental implants, metallic cranial plates, screws, burr hole covers, bone substitute device, magnetic metallic implants/electrodes/valves placed in upper limbs , torso, or higher, etc. If you have any questions regarding the implant, consult your physician or the manufacturer of your implant.

2- Updated Warning:
Magnets are used in the mask and headgear clips with a field strength of 380mT. With the exception of the devices identified in the contraindication, ensure that the mask is kept at least 6 inches (approx. 16 cm) away from any other medical implants or medical devices that can be impacted by the magnetic fields to avoid possible effects from localized magnetic fields. This applies to you or anyone in close physical contact with your mask.

If a patient, or anyone in close physical contact, has an active medical implant or metallic implant, they should contact their mask supplier to find a replacement mask that does not include magnets. If the patient is unsure whether or not they should use the mask, they should consult their physician or the manufacturer of their implant.

In case you experience harm from this product, it is important to report it. It can help to detect & resolve issues and prevent others from being harmed, and it enables better surveillance. If symptoms persist, seek medical care.

Company name: Sleepnet Corporation
Brand name: Mojo, Mojo 2, iQ 2, Phantom 2
Product recalled: CPAP and BIPAP Masks with Magnets
Reason of the recall: Potential interference with certain medical implants
FDA Recall date: March 18, 2024

Source: www.fda.gov/safety/recalls-market-withdrawals-safety-alerts/sleepnet-corporation-issues-worldwide-recall-cpap-and-bipap-masks-magnets-due-potential-interference

#medicaldevices #us

Recent Interesting Reports

G
Saw an advertisement on Facebook from Amazon warehouse that said pallets of returned electronics were being sold for $29. They needed to empty the warehouse. I bought one only to receive a small mystery box with a watch inside. I thought I had ordered it wrong so… See more

#scam #delivery #facebookscam #middleton #wisconsin #us

T
I ordered some clothes that I still haven't received yet but I received a ring instead. It says to scan this QR code to register it. I thankfully didn't do that because I wonder if that somehow takes my money? It came from Angelo Want

#scam #onlineshopping #onlinescam #delivery #5549arrowhighway #montclair #california #us

Last 30 days