Report by
Safety Report: Cardinal Health’s Monoject™ Prefilled Syringes - recalled due to potential air draw back, USA
3 years ago •source fda.gov
Recall notice
United States
On August 19, 2021, Aligned Medical Solutions initiated a nationwide recall of Cardinal Health’s Monoject™ Flush Prefilled Saline Syringes placed into 9,378 kits. Including:- 1 lot of AMS-9041CP Leaderflex Insertion Kit with Ultrasound
- 1 lot of AMS-9046CP-1 Insertion Tray-RX
- 45 lots of AMS8939A Universal Procedure Pack w/Split Drape
- 1 lot of AMS9957A Port Insertion Pack
- 3 lots of AMS12149 Procedure Pack
These convenience kits have been found to contain the Cardinal Health’s Monoject™ Flush Prefilled Saline Syringe part # 8881570121, which has been recalled for the potential for the plunger to draw back after the air has been expelled and reintroduced air back into the syringe. If a clinician is not aware of air being introduced into the syringe, the clinician could inadvertently push air into the vascular system creating the potential for an air embolism. Injection of air into the vascular system can cause air embolism which can result in serious adverse health consequences or death.
Consumers who have affected product(s) should immediately review their inventory and quarantine all affected kits. Customers will be provided with a yellow label that is to be placed on the packs containing the recalled Cardinal Health’s Monoject™ Flush Prefilled Saline Syringe. The label will read as follows; At the time the kit is opened for use the prefilled syringe manufactured by Cardinal Health should be identified, set aside and not used. Do not use the prefilled syringe in the kit.
The recalled prefilled syringe should be rendered unusable to protect against inadvertent use and disposed or pursuant to the medical waste policies in effect at your institution
Recalled Product(s) were manufactured from 01/13/2020 to 10/14/2020 and distributed from 01/23/2020 to 10/19/2020. The following kit part numbers and lot numbers containing the recalled /affected Cardinal Health Monoject™ Flush Prefilled Syringes have been recalled:
- Formula No. AMS-9041CP. Description: 136833. Item No.: Leaderflex Insertion Kit with Ultrasound.
- Formula No. AMS-9046CP-1. Description: 137794. Item No.: Insertion Tray-RX.
- Formula No. AMS8939A . Description: 140390, 141087, 141088, 141096, 141097, 141098, 141100, 141101, 141102, 141717, 145615, 146037,146039, 146349, 146351, 146766, 146768, 146905, 146908, 147902, 147903, 148409, 148410, 148411, 148412, 148413, 148414, 148450, 148451, 148452, 148636, 148638, 148639, 148640, 149131, 149132, 149133, 149134, 149135, 149136, 149137, 149138, 149484, 149485, 149486. Item No.: Universal Procedure Pack w/Split Drape.
- Formula No. AMS9957A. Description: 147773. Item No.: Port Insertion Pack.
- Formula No. AMS12149. Description: 145718, 146897, 148612. Item No.: Procedure Pack.
Product(s) can be identified by the pack label inside the sterile barrier of the convenience kit.
Aligned Medical Solutions has notified the FDA of this action.
No injuries have been reported to date.
Aligned Medical Solutions is notifying its distributors by email and will arrange for labels to identify the packs with the recalled syringe. Aligned Medical Solutions will also arrange for replacement syringes to be sent for all recalled product(s).
Aligned Medical Solutions distributed these packs Nationwide.
This recall is being conducted with the knowledge of the U.S. Food and Drug Administration.
Company name: Windstone Medical Packaging dba Aligned Medical Solutions
Brand name: Cardinal Health’s Monoject™
Product recalled: Flush Prefilled Saline Syringes
Reason of the recall: Potential for the plunger to draw back after the air has been expelled and reintroduced air back into the syringe
FDA Recall date: October 27, 2021
Check the full recall details on www.fda.gov/safety/recalls-market-withdrawals-safety-alerts/windstone-medical-packaging-dba-aligned-medical-solutions-issues-nationwide-recall-cardinal-healths
Source: FDA