Infusion System Administration Set (infusion set) - recalled due to Use of the pump system potentially could cause over-infusion or under-infusion of therapy, USA

3 years ago source www.fda.gov

Recall notice

United States

Company name: CME America
Brand name: BodyGuard
Product recalled: Infusion System Administration Set (infusion set)
Reason of the recall: Use of the pump system potentially could cause over-infusion or under-infusion of therapy
FDA Recall date: July 08, 2020
Recall details: PRNewswire/ - CME America, a wholly owned subsidiary of BD (Becton, Dickinson and Company), announced a voluntary recall for all CME America BodyGuard® Infusion System Administration Sets (infusion sets) used with the company's BodyGuard® infusion pumps—that were distributed beginning May 2016 (see complete impacted product list below).  This action was initiated on June 16, 2020. As part of CME America's commitment to quality, following the previously announced recall (April 27, 2020) of the BodyGuard® Infusion Pump Systems, the company conducted additional flow-rate accuracy testing.

This testing revealed that some infusion sets do not meet the ±5% delivery accuracy for the system or the ±13% accuracy identified in the earlier recall notification (www.bd.com Therefore, the use of the pump system potentially could cause over-infusion or under-infusion of therapy and patient harm. Based on those test results CME America is providing additional information and customer actions regarding its previous recall to include all infusion sets used with the BodyGuard® infusion pump distributed beginning May 2016.

CME America has defined four categories of impacted infusion sets based on delivery inaccuracy variability, two of which (Category A and B) can continue to be used in accordance with the instructions in the recall letter (www.bd.com CME America also announced on April 27, 2020, the decision to suspend distribution of the BodyGuard® infusion pumps and to remove all existing products from the U.S. market. CME America will work with customers to address the latest expanded infusion set recall and will continue to maintain continuity of care during the COVID-19 pandemic. Until such time that the BodyGuard® pumps have been removed from the market, CMEA will supply accessories and infusion sets in "Category A" and "Category B" to support the infusion pumps, and customers can continue to use the products in accordance with the Operator's Manual and the additional mitigations outlined in the customer letter.

Affected Product Types:

- BodyGuard BodySet
- BodyGuard Microset with Needleless Adaptor
- BodyGuard Microsets
- BodyGuard Microset w/ Non-Vented Spike Connector
- BodyGuard set with Female Luer
- BodyGuard Microset w/ 0.2 mic filter and lower y-site with female luer
- BodyGuard Microsets with Filter
- Standard BodySet with Needleless Connectors
- BodyGuard Microset with Filter and Manual Priming Valve
- CMExpress Microbore Sets
- BodyGuard Microset with Male Luer Connectors
- CMExpress Needleless Y Site Microbore Set

Check the full recall details on www.fda.gov

Source: FDA

#medicaldevices #us

Recent Interesting Reports

Q

No Package, Durham, North Carolina, USA

3 days ago reported by user-qcchx737

I ordered clothes from an Instagram add almost a month ago. The “tracking” given shows limited information and according to the tracking website it was delivered however I have yet to receive the package. The website has no phone number only an email address. I got a … See More
response from service@ risktic. com the first time when I emailed them a day after the package was said to have been delivered and I was instructed to wait 48 hours. I waited then sent another email (the only source of communication) to the same address but this time no response. I sent another email asking for a complete refund, the prior email I only requested the dresses. The response from this email was a different address support@ lixvb. com and asked if they could refund 50%. This is a complete scam. Don’t buy from them. It’s been 30 day and still no dresses.

https: //risktic. com
See Less

#onlineshopping #scam #onlinescam #delivery #durham #northcarolina #us

Create a Report

Please provide report title
Please provide location
Please provide details

Get alerts and updates for your case!

This information is not shown on the website. We can notify you of updates, and may contact you for more information to help resolve or follow up on your issue.

Please provide email or phone
Please provide email or phone

We do not show your email or contact info

Skip this step

Add photos or video

We recommend photos and videos to help explain your report

Add Photo/Video
Skip this step

Email or SMS copy of report

Enter below to get emailed a copy of your report, or sms a link to your report

Please provide email or phone
Please provide email or phone

Thank you for your report!


IMPORTANT - Your report is queued
It may take up to 12 hours to process your report.


Email us at support@safelyhq.com if you have questions.

Report by

Please provide location
Please provide location
Please provide report title
Please provide details
Please accept our Terms & Conditions
Get alerts and updates for your case! Your contact information is not shown on the website.

This information is not shown on the website. We can notify you of updates, and may contact you for more information to help resolve or follow up on your issue.

or
Please provide email or phone
Thank you for your report!

IMPORTANT - YOUR REPORT IS QUEUED - IT MAY TAKE UP TO 12 HOURS FOR YOUR REPORT TO SHOW ON OUR HOME PAGE (IF NOT OPTED AS PRIVATE)

Visit our learn pages for more helpful information or, email us: support@safelyhq.com

Goodbye! and best wishes in your recovery
Last 30 days